Model | Se-240R (Type 1) | Se-240T (Type 2) |
Internal Size WDH | 1100 ×550 ×400mm | 400 ×1170 ×500mm |
External Size WDH | 1340 ×1250 ×1560mm | 550 ×1300 ×1480mm |
Temperature Range | RT+10℃~+70℃ | RT+10℃~+80℃ |
Blackboard Temp Range | +50℃~70℃ | 80℃ |
Temp Fluctuation | ≤0.5℃ | ≤0.5℃ |
Temp Uniformity | ≤2.0℃ | ≤2.0℃ |
Functions | UV Irradiation + Condensation + Water Spray | UV Irradiation + Condensation + Water Spray |
Shelf Quantity | Tray shelf×1 | Plate shelf×24 |
Shelf Material | Stainless steel | Aluminum alloy |
Sample Requirement | No requirement | 75×290mm LW Size |
UV Lamp Irradiance | 0 ~ 2.0w/m2 | 0 ~ 1.0w/m2 |
Irradiance Range | 1100 ×550 ×400mm | 1020×290mm |
UV Lamp Wavelength | 280—400nm | 315—400nm |
UV Lamp Type | UVA340, UVB313-Option | UVA340, UVB313-Option |
UV Lamp Quantity | 8 pcs | 8 pcs |
UV Lamp Length | 1200mm | 1200mm |
UV Lamp Power | 40w each | 40w each |
UV Lamp Life | 1600 h | 1600 h |
UV Lamp Location | At the top of working room | At the both sides of tower structure, each side is 4 pcs |
Distance Between Sample and Lamp | 400mm (adjustable) | 50±3mm |
Spray Nozzle Diameter | Ф0.8 mm | Ф0.8 mm |
Test Period | 1 ~ 9999 h | 1 ~ 999 h |
Internal Material | SUS304 mirror stainless steel | SUS304 mirror stainless steel |
External Material | Cold rolled steel plate white electrostatic spraying | SUS304 stainless steel |
Controller | 7.0 inch program touch screen controller, RS232/485 port, USB data port + English computer software |
Protection | Over-temperature protection, Over-temperature of compressor protection, Power phases protection. |
Cooling Way | Air cooling |
Sensor | Temperature: PT100 |
Power Supply | 220 ~ 240V, 50Hz, 1-phase or 380 ~ 415V, 50Hz, 3-phases |
Warranty | 1 year |
Standards | Wheel caster×4, User manual×1, Controller manual×1 |